Climatic Temperature & Humidity testing
EMC Technologies have comprehensive Environmental Laboratories offering NATA & A2LA accredited Temperature and Humidity cycling test services
EMC Technologies environmental testing
Introduction
High/Low-Temperature test is to evaluate material likely to be deployed in areas where temperatures (ambient or induced) are higher/lower than standard ambient.
Humidity test is to evaluate material that is likely to be stored or deployed in a warm, humid environment, an environment in which high levels of humidity occur, or to provide an indication of potential problems associated with humidity.
EMC Technologies is A2LA and NATA accredited for temperature and humidity environmental testing.
We have large programmable Environmental Chambers for Temperature and IP testing of dust & water ingress protection at both Melbourne and Sydney branches which can accommodate most consumer products. These chambers are often utilised for various EMC/Radio/Safety standards which specify environmental tests such as AS/NZS 4268, AS/NZS 60950.1, EN 300 328, EN 300 220, FCC Part 15C and RSS-210.
Melbourne – Equipment specifications:
- Temperature range: -70°C ~ +180°C
- Internal chamber size: 1100x950x950(mm)
- Rate of cooling: 2.5°C/Minute (Note1)
- Rate of heating: 4°C/Minute (Note2)
- Humidity range: 10% RH to 98% RH
Note1 & 2: with an empty chamber in accordance with IEC60068-3-5
Sydney – Equipment specifications:
- Capacity: 900 x 900 x 800 Width x Height x Depth (mm)
- Temperature range: -20 ~ +95C with an accuracy of ±2%
- Humidity range: 30 ~ 95% with an accuracy of R.H. ±2%
Environmental Testing – High/Low Temperature Testing
Our A2LA accreditation covers (pending for some standards) the following standards
MIL-STD-810D Method 501.2, 502.2
MIL-STD-810E Method 501.3, 502.3
MIL-STD-810F Method 501.4, 502.4
MIL-STD-810G Method 501.6, 502.6
MIL-STD-810H Method 501.7, 502.7
MIL-STD-202H Method 108
RTCA/DO-160G Section 4
IEC 60068-2-1
IEC 60068-2-2
IEC/EN 50155 (Table 12, Test 5, 6, 7, 8)
AS4428.6 (Clause 17.4, 17.5, 17.10, 17.12)
IEC/EN 60945 (Clause 8.2, 8.3, 8.4)
EN 302 054
Environmental Testing – Humidity Testing
Our A2LA accreditation (pending for some standards) covers the following standards